- Tytuł:
- Prototype Silicon Position-Sensitive Detector Working with Bragg-Brentano Powder Diffractometer
- Autorzy:
-
Zięba, A.
Dąbrowski, W.
Gryboś, P.
Powroźnik, W.
Stobiecki, T.
Świentek, K.
Słowik, J.
Wiącek, P. - Powiązania:
- https://bibliotekanauki.pl/articles/2030562.pdf
- Data publikacji:
- 2002-05
- Wydawca:
- Polska Akademia Nauk. Instytut Fizyki PAN
- Tematy:
-
07.85.-m
29.40.Wk - Opis:
- A prototype 64-channel detector module, comprising a silicon strip detector with strip pitch of 100μm and 64-channel ASIC RX64, was tested with the X-Pert Philips MPD diffractometer. Basic parameters of the detector module, energy resolution, and detection efficiency, were evaluated as a function of the counting rate. Energy resolution of 1.1 keV FWHM for photon rate up to 1×10$\text{}^{7}$ photon/s per 1 cm of the active width of the detector was demonstrated. The prototype detector, when applied in a diffractometer utilizing Bragg-Brentano focusing principle, allows to increase the counting rate by about 2 orders of magnitude with respect to a single counter. Exemplary diffraction patterns of polycrystalline samples of Si and SiO$\text{}_{2}$ (quartz peak cluster) are presented.
- Źródło:
-
Acta Physica Polonica A; 2002, 101, 5; 629-634
0587-4246
1898-794X - Pojawia się w:
- Acta Physica Polonica A
- Dostawca treści:
- Biblioteka Nauki