- Tytuł:
-
Automation measurement of intra-abdominal pressure
Automatyzacja pomiarów ciśnienia wewnątrzbrzusznego - Autorzy:
-
Jurčišin, M.
Cabúk, P.
Slosarčík, S. - Powiązania:
- https://bibliotekanauki.pl/articles/158933.pdf
- Data publikacji:
- 2011
- Wydawca:
- Sieć Badawcza Łukasiewicz - Instytut Elektrotechniki
- Tematy:
-
ciśnienie wewnątrzbrzuszne
pomiar ciśnienia
pojemnosciowe czujniki ciśnienia
metoda nieinwazyjna
zastosowanie kliniczne
automation
measurements
capacitive pressure sensors
calibration
intra-abdominal pressure - Opis:
-
The article deals with the design of capacitive pressure sensor for automated measurement of intra-abdominal pressure by a non-invasive method. A capacitive pressure sensor consists of a glass tube with evaporated thin film electrodes. The capacity change is given by the level change of displaced fluid in the tube. For capacitive sensors with different inner diameter of tubes calibration curves were experimentally found. A capacitive sensor is a subpart of an automated measuring system for clinical application of intra-abdominal pressure measurement.
Artykuł dotyczy budowy pojemnościowych czujników ciśnienia do automatycznego pomiaru ciśnienia wewnątrzbrzusznego metodą nieinwazyjną. Pojemnościowy czujnik cieśnienia składa się ze szklanej rurki z naparowanymi cienkowarstwowymi elektrodami. Zmiana pojemności objawia się przez zmianę poziomu przemieszczonej cieczy w rurce. Wyznaczono doświadczalnie krzywe wzorcowania dla czujników ciśnienia z różnymi średnicami wewnętrznymi rurek. Czujnik pojemnościowy jest podzespołem automatycznego układu pomiarowego w klinicznym zastosowaniu pomiaru ciśnienia wewnątrzbrzusznego. - Źródło:
-
Prace Instytutu Elektrotechniki; 2011, 252; 93-100