- Tytuł:
- Validation of transient time method to calibrate oil flow meters in closed conduits using 123I as the radiotracer
- Autorzy:
-
Gonçalves, Eduardo R.
Brandão, Luís E. B.
Braz, Delson - Powiązania:
- https://bibliotekanauki.pl/articles/2055689.pdf
- Data publikacji:
- 2021
- Wydawca:
- Instytut Chemii i Techniki Jądrowej
- Tematy:
-
calibration flow meters
oil flow rates
iodine-123
transient time method - Opis:
- To assure the credibility of commodity transfer operations in the oil and gas industry, calibrated flow meters are used to quantify the movement of fluids in the pipelines. The purpose of this paper is to validate the transient time method (TTM) to calibrate oil flow meters installed in restricted areas, using the 123I-labelled oil as a radiotracer. Traditionally, as proposed by the standard ISO 2975-7:1977 [1] for experiments in an aqueous medium, the TTM is employed by positioning two detectors at separate locations. However, in industrial plants, it is not always possible to install detectors at the distances recommended by the ISO 2975-7. The method proposed in this paper uses four scintillator detectors separated one from each other by 0.30 m and three injections containing 5.0 ml of 123I-labelled oil. The experiments were carried out in an oil flow rig with a turbulent flow profile. The results have reached an uncertainty which is lower than 1.0%.
- Źródło:
-
Nukleonika; 2021, 66, 2; 69--74
0029-5922
1508-5791 - Pojawia się w: