- Tytuł:
- Finite element analysis of the percutaneous coronary intervention in a coronary bifurcation
- Autorzy:
-
Bukała, J. K.
Małachowski, J.
Kwiatkowski, P. - Powiązania:
- https://bibliotekanauki.pl/articles/307501.pdf
- Data publikacji:
- 2014
- Wydawca:
- Politechnika Wrocławska. Oficyna Wydawnicza Politechniki Wrocławskiej
- Tematy:
-
angioplastyka
FEA
symulacja
angioplasty
balloon
kissing
simulation - Opis:
- Purpose: The paper presents the process of numerical modeling and simulation of balloon angioplasty of the coronary artery using Finite Element Method. Methods: The authors focused on the issue of adequate pressure applying in an arterial tissue during the post-dilatation process in the 'kissing balloon' stenting technique applied to patients with bifurcation stenosis. Despite great progress in the field of interventional cardiology, angioplasty in bifurcations still belongs to the most difficult interventions, generally being less effective and more risky than in the cases of straight stenosis.Results: During the modeling procedures and further simulations, the authors focused on mutual cooperation of non - compliant angioplasty balloons and the coronary artery. The other goal was to develop a sufficiently accurate model of the coronary artery fragment, including its bifurcation and angioplasty balloons, however it was decided to ignore the modeling of coronary stents.Conclusions: The undertaken issue is considered as relatively complicated and complex but, in the authors opinion, the implementation of advanced computer aided engineering techniques may, in this case, answer several important questions without the need of performing costly and aggravating in vivo tests.
- Źródło:
-
Acta of Bioengineering and Biomechanics; 2014, 16, 4; 23-31
1509-409X
2450-6303 - Pojawia się w:
- Acta of Bioengineering and Biomechanics
- Dostawca treści:
- Biblioteka Nauki