Tytuł:
Estimation of quantization steps of time interpolator in view of temperature changes
Autorzy:
Sondej, D.
Szplet, R.
Powiązania:
https://bibliotekanauki.pl/articles/114194.pdf
Data publikacji:
2017
Wydawca:
Stowarzyszenie Inżynierów i Techników Mechaników Polskich
Tematy:
time-digital conversion
two-stage interpolation
multi-edge coding in independent coding lines
time counter
time digitizer
Opis:
This paper presents a method of limiting the effect of ambient temperature drift on the measurement uncertainty of a time counter (TC). A change in ambient temperature causes a change in the TC transfer function, i.e. the widths of quantization steps to be exact. Recalibration is a procedure that is then required, but it disturbs the measurement process. However, with the knowledge of the current ambient temperature and having the set of transfer functions identified at different temperatures, it is possible to determine and use the most adequate transfer function and virtually eliminate the temperature impact. For this purpose, three interpolation methods were studied: the nearest neighbor method, linear and polynomial interpolations. A newly evaluated transfer functions were tested in interpolating TC to select the best interpolation method.
Źródło:
Measurement Automation Monitoring; 2017, 63, 5; 192-194
2450-2855
Pojawia się w:
Measurement Automation Monitoring
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
A study of the effect of temperature changes on the interpolating time counter
Autorzy:
Sondej, D.
Szplet, R.
Powiązania:
https://bibliotekanauki.pl/articles/114363.pdf
Data publikacji:
2015
Wydawca:
Stowarzyszenie Inżynierów i Techników Mechaników Polskich
Tematy:
time interval measurement
time-to-digital conversion
two-stage interpolation
time counter
Opis:
This paper presents an analysis of the impact of ambient temperature changes on main parameters of the interpolating time counter. The performed tests reveal that a relatively small change in the ambient temperature of 1°C causes a measurement error of the counter as large as 3.5 ps. The thorough research of two stages of interpolation of the counter allowed determining the main sources of the error. One of them is the temperature drift of widths of four-phase clock (FPC) segments in the first interpolation stage (FIS). It equals 2.5 ps/°C. The widths of FPC phases directly influence the active range of the second interpolation stage (SIS) and its offset. The test results also show that the temperature drift of the offset has a greater impact on the measurement accuracy than the temperature-driven changes of quantization steps in SIS. The presented conclusions are the first step to develop a new method for reducing the impact of changes in the ambient temperature on the measurement accuracy of the interpolating time counter.
Źródło:
Measurement Automation Monitoring; 2015, 61, 7; 302-304
2450-2855
Pojawia się w:
Measurement Automation Monitoring
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
Bezpośredni przetwornik czas-liczba z kodowaniem wielokrotnym
A direct time-to-digital converter with multiple coding
Autorzy:
Szplet, R.
Sondej, D.
Grzęda, G.
Powiązania:
https://bibliotekanauki.pl/articles/151174.pdf
Data publikacji:
2013
Wydawca:
Stowarzyszenie Inżynierów i Techników Mechaników Polskich
Tematy:
precyzyjna metrologia czasu
przetwornik czas-liczba
metoda kodowania wielokrotnego
układy FPGA
precise time metrology
time-to-digital converter
wave union method
FPGA devices
Opis:
W artykule opisane są projekt i wyniki badań przetwornika czas-liczba o rozdzielczości 5,3 ps (1 LSB) i zakresie pomiarowym 428 ps. Do przetwarzania czasowo-cyfrowego użyta została metoda kodowania wielokrotnego. Metoda ta umożliwia pokonanie ograniczeń technologicznych współczesnych układów scalonych i uzyskanie wartości rozdzielczości mniejszej niż czas propagacji pojedynczej komórki linii kodującej. Przetwornik został zrealizowany w układzie programowalnym Spartan-6 firmy Xilinx.
This paper presents the implementation and tests results of a time-to-digital converter based on the wave union method and implemented in Spartan-6 FPGA (Xilinx). The converter has the resolution of 5,3 ps (1 LSB) in the measurement range of 428 ps and the integral nonlinearity of 3,8 LSB (Fig. 7). In the wave union method, contrary to the typical conversion methods with a single coding, the resolution is lower than the FPGA cell delay thanks to coding several transitions of the time event signal (Fig. 2). In addition, the linearity of conversion is increased by reducing the width of wide bins. Although, using a multi-transition pattern gives better performance, it also brings more problems to be solved. The main problems such as implementation of a pattern generator for certain amount of transitions, minimal delays between transitions and elimination of bubble errors are discussed in this paper. The pattern generator (Fig. 3) is implemented with use of a carry chain. It enables controlling the pattern by means of diagnostic and measurement software. Bubble errors (Fig. 4) are eliminated with a fast asynchronous encoder (Fig. 5). The diagnostic-control software (Fig. 6) allows to configure the pattern generator, launch the measurement session and generate a text file with all information needed to calculate conversion characteristics of the time-to-digital converter.
Źródło:
Pomiary Automatyka Kontrola; 2013, R. 59, nr 8, 8; 842-844
0032-4140
Pojawia się w:
Pomiary Automatyka Kontrola
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
Evaluation of Selected Timing Parameters of FPGA Device
Autorzy:
Sondej, D.
Szymanowski, R.
Szplet, R.
Powiązania:
https://bibliotekanauki.pl/articles/114158.pdf
Data publikacji:
2018
Wydawca:
Stowarzyszenie Inżynierów i Techników Mechaników Polskich
Tematy:
propagation time
jitter
ring oscillator
delay line
carry chain
LUT
FPGA
Opis:
We present a method and results of measurements of FPGA (Field Programmable Gate Array) selected timing parameters crucial in many timing sensitive applications such as precise time and frequency metrology. Two main parameters, i.e. the delay and its jitter, were evaluated for look-up-tables (delay 740 ps/jitter 1.33 ps), IO buffers (na/0.45 ps) and carry-chain multiplexers (28ps/0.153 ps) integrated in a programmable device Spartan-6 (Xilinx) which is one of most popular FPGA chips on the market now. Measurements were performed with the use of fast real-time sampling oscilloscope.
Źródło:
Measurement Automation Monitoring; 2018, 64, 1; 23-25
2450-2855
Pojawia się w:
Measurement Automation Monitoring
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
An autonomous microcontroller system for controlling a multi-channel time counter
Autorzy:
Sondej, D.
Sawicki, M.
Szplet, R.
Powiązania:
https://bibliotekanauki.pl/articles/114671.pdf
Data publikacji:
2015
Wydawca:
Stowarzyszenie Inżynierów i Techników Mechaników Polskich
Tematy:
microcontroller
time counter
autonomous system
time-to-digital converter
Opis:
The paper presents the design of a microprocessor system intended for control, data processing and communication in a multi-channel time counter. The time counter is a relatively complex autonomous instrument designed for measurements of time with picosecond precision and frequency within a range of 3.5 GHz. The device employs a high-speed, single-chip microcontroller with event-driven programming without the mediation of an operating system. The autonomous operation of the measurement system with real-time controlling and data processing was achieved. The paper focuses on the hardware design and the software development model which enables collision-free and concurrent work of many components. The device uses advanced mechanisms available in STM32 series microcontrollers, allowing the efficient support for USB and Ethernet interfaces. The microprocessor system works at a relatively low frequency, which minimizes emission of interference and allows measuring time intervals with a high precision.
Źródło:
Measurement Automation Monitoring; 2015, 61, 7; 305-307
2450-2855
Pojawia się w:
Measurement Automation Monitoring
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
Oprogramowanie diagnostyczno-sterujące interpolacyjnego licznika czasu w układzie programowalnym
Diagnostic and control software for the interpolating time counter implemented in a programmable device
Autorzy:
Grzęda, G.
Sondej, D.
Szplet, R.
Powiązania:
https://bibliotekanauki.pl/articles/155302.pdf
Data publikacji:
2014
Wydawca:
Stowarzyszenie Inżynierów i Techników Mechaników Polskich
Tematy:
precyzyjna metrologia czasu
oprogramowanie diagnostyczno-sterujące
kodowanie wielokrotne w niezależnych liniach kodujących
model warstwowy
precise time metrology
diagnostic-control software
multiedge coding in independent coding lines
layer model
Opis:
W artykule opisano projekt oprogramowania diagnostyczno-sterującego licznika czasu z kodowaniem wielokrotnym w niezależnych liniach kodujących, wykonanego w układzie programowalnym Spartan-6 firmy Xilinx. Przedstawiono sposób sterowania licznikiem czasu, koncepcję oprogramowania sterującego, jego zadania oraz warstwową budowę. Opisano graficzny interfejs użytkownika programu i jego funkcjonalność. Prezentowane są także wyniki badań eksperymentalnych licznika czasu.
This paper presents the diagnostic and control software of a time interval counter with multi-edge coding in independent coding lines, implemented in the Spartan-6 FPGA device manufactured by Xilinx. The method of time-to-digital conversion [1] is presented along with the design of the time interval counter (Fig. 1). Subsequently, the main goals of the control software, along with its logical structure, are described. The paper shows the layer model (Fig. 2) of the program, reveals the method of communication with the time counter and the way of decoding measurement frames. The bottom-most communication layer transfers the data through USB to the device. The next control layer operates on hardware registers and the measurement layer calibrates the counter and triggers measurements. Finally, the graphic user interface (GUI) layer binds the application together and steers the user interface. The program operates in two main modes: calibration and time interval measurement. Apart from both these modes, the data flow across the layers and the way of saving data generated during counter operation are described. The GUI (Fig. 3) is described as well, showing the main types of operation along with the capabilities of configuring the calibration and measurement processes. Finally, the paper presents the test results of the time counter in both main operation modes (Fig. 4).
Źródło:
Pomiary Automatyka Kontrola; 2014, R. 60, nr 7, 7; 441-443
0032-4140
Pojawia się w:
Pomiary Automatyka Kontrola
Dostawca treści:
Biblioteka Nauki
Artykuł
    Wyświetlanie 1-6 z 6

    Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies