- Variation of Energy Resolution with Distance for a NaI(Tl) Detector
- Autorzy:
-
Akkurt, İ.
Arda, S.
Gunoglu, K. - Powiązania:
- https://bibliotekanauki.pl/articles/1402489.pdf
- Data publikacji:
- 2015-08
- Wydawca:
- Polska Akademia Nauk. Instytut Fizyki PAN
- Tematy:
-
29.40.-n
29.30.Kv - Opis:
- As the nuclear radiation has started to be used in a variety of different fields, it is important to be protected from it, and thus the radiation measurement becomes vital. The quality of the performance of a detection system, used for the energy measurements, is important. It is characterized by the width of the pulse-height distribution, obtained for the particles of the same energy (monoenergetic source). The energy spectrum of a radiation source depends on the type and energy of the incident particle and the type of the detector. In this work the energy resolution of a 3"×3" NaI(Tl) detector has been measured for photon energies of 511, 662, 835, 1173, 1275, and 1332 keV, and its variation with the detector-source distance was investigated. The energy resolution of a detector system is obtained from the peak full width at one-half of the maximum height (FWHM) of a single peak (for a particular energy) as a function of detector-source distance. It was found that the energy resolution has decreased with the increasing distance.
- Źródło:
-
Acta Physica Polonica A; 2015, 128, 2B; B-422-B-423
0587-4246