Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Wyszukujesz frazę "noise uncertainty" wg kryterium: Temat


Wyświetlanie 1-2 z 2
Tytuł:
Investigation of noise in surface topography measurement using structured illumination microscopy
Autorzy:
Li, Zhen
Gröger, Sophie
Powiązania:
https://bibliotekanauki.pl/articles/2052124.pdf
Data publikacji:
2021
Wydawca:
Polska Akademia Nauk. Czytelnia Czasopism PAN
Tematy:
surface topography measurement
measurement noise
uncertainty
structured illumination microscopy
Opis:
Noise is a fundamental metrological characteristic of the instrument in surface topography measurement. Therefore, measurement noise should be thoroughly studied in practical measurement to understand instrument performance and optimize measurement strategy. This paper investigates the measurement noise at different measurement settings using structured illumination microscopy. The investigation shows that the measurement noise may scatter significantly among different measurement settings. Eliminating sample tilt, selecting low vertical scanning interval and high exposure time is helpful to reduce the measurement noise. In order to estimate the influence of noise on the measurement, an approach based on metrological characteristics is proposed. The paper provides a practical guide to understanding measurement noise in a wide range of applications.
Źródło:
Metrology and Measurement Systems; 2021, 28, 4; 767-779
0860-8229
Pojawia się w:
Metrology and Measurement Systems
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
Evaluation of a commercial high resistance bridge and methods to improve its precision
Autorzy:
Mihai, Iulian
Capra, Pier Paolo
Galliana, Flavio
Powiązania:
https://bibliotekanauki.pl/articles/2173898.pdf
Data publikacji:
2022
Wydawca:
Polska Akademia Nauk. Czytelnia Czasopism PAN
Tematy:
high resistance measurements
dual source bridge
DSB
settle-time
measurement uncertainty
compatibility test
measurement noise
Opis:
At the National Institute of Metrological Research (INRIM) an evaluation of a commercial dual source high resistance bridge has been performed. Its two main measurement modes (single measurements and multiple measurements) have been investigated. The best settle time of a 10:1 measurement of high resistance ratio has been estimated to be about three times the time constant of the circuit involving the resistors. This constant, in turn, depends on the highest value resistor. By means of mathematical estimators, suitable numbers of the readings of the detector have been established in order to minimize noises. A compatibility test at 100 TΩ has shown that the best precision of the commercial bridge is achieved utilizing the multiple measurements mode with the auto update function. This mode also allows the characterization of a resistor as a function of the settle time. This characterization can be useful for the owner of the resistor who can request the laboratory to perform the calibration of the resistor with the settle time which is necessary for him.
Źródło:
Metrology and Measurement Systems; 2022, 29, 4; 701--718
0860-8229
Pojawia się w:
Metrology and Measurement Systems
Dostawca treści:
Biblioteka Nauki
Artykuł
    Wyświetlanie 1-2 z 2

    Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies