Measurement and analysis of the air pressure curve on the rigid lenses by use of ocular response analyzer
Autorzy:
Kasprzak, H.
Mazur, E.
Widlicka, M.
Powiązania:
https://bibliotekanauki.pl/articles/307438.pdf
Data publikacji:
2014
Wydawca:
Politechnika Wrocławska. Oficyna Wydawnicza Politechniki Wrocławskiej
Tematy:
okulistyka
ciśnienie śródgałkowe
lepkosprężystość
ORA
ocular viscoelasticity
intraocular pressure
air pressure curve
Opis:
Purpose: Pressure curves obtained from Ocular Response Analyzer (ORA) differ for the same patient in form and height. In some cases measurements on the subject show significant differences between recorded pressure curves. The purpose of the paper is to examine if the differences result from the corneal properties or from the device operation. Methods: Examination of air pressure curves was carried out on four plano-convex glass lenses with radii of curvature close to the central corneal radius. Lenses were mounted in front of the air jet of the ORA analyzer. Series of 30 measurements on each lenses were recorded with 20s and 60s time interval between measurements. Results were exported to computer and analyzed numerically Results: Results show much higher reproducibility of pressure curves in every series of measurements in comparison to pressure curves recorded on the patients eye. This demonstrates that ORA produces air pulses with high reproducibility. Conclusions: Differences in air pressure pulses for the real eye can indicate the dynamics of ocular properties during measurements. Obtained pressure curves are not symmetrical and not well fitted by Gaussian curve. Type of asymmetry of air pressure curves may be explained by viscoelasticity of air.
Źródło:
Acta of Bioengineering and Biomechanics; 2014, 16, 4; 117-121
1509-409X
2450-6303
Pojawia się w:
Acta of Bioengineering and Biomechanics
Dostawca treści:
Biblioteka Nauki
Artykuł
    Wyświetlanie 1-1 z 1

    Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies