Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Wyszukujesz frazę "Cardiac output" wg kryterium: Temat


Wyświetlanie 1-2 z 2
Tytuł:
Theoretical and model analysis of the unreliability of cardiac output measurement by means of the thermodilution method
Autorzy:
Gawlikowski, M.
Pustelny, T.
Przywara-Chowaniec, B.
Nowak-Gawlikowska, J.
Powiązania:
https://bibliotekanauki.pl/articles/200095.pdf
Data publikacji:
2011
Wydawca:
Polska Akademia Nauk. Czytelnia Czasopism PAN
Tematy:
cardiac output measurement
thermodilution method
Opis:
Thermodilution is the clinically most often applied method of cardiac output measurements. This method is based on thermal indicator (iced isotonic salt solution) variation measurements by a Swan-Ganz catheter located inside the pulmonary artery. The unreliability of thermodilution should be estimated theoretically because of the lack of references. In this paper an attempt has been made to estimate theoretically the unreliability of thermodilution cardiac output measurements.
Źródło:
Bulletin of the Polish Academy of Sciences. Technical Sciences; 2011, 59, 4; 435-439
0239-7528
Pojawia się w:
Bulletin of the Polish Academy of Sciences. Technical Sciences
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
The influence of limited dynamic response of the indicator detector in a Swan-Ganz catheter on the overestimation of cardiac output measurement by means of thermodilution
Autorzy:
Gawlikowski, M.
Pustelny, T.
Powiązania:
https://bibliotekanauki.pl/articles/221141.pdf
Data publikacji:
2012
Wydawca:
Polska Akademia Nauk. Czytelnia Czasopism PAN
Tematy:
cardiac output measurement
thermodilution method
Swan-Ganz catheter
Opis:
Nowadays the "gold clinical standard" of hemodynamics diagnostic and cardiac output measurements is pulmonary artery catheterization by means of the Swan-Ganz catheter and thermodilution. The method itself is sensitive to numerous disturbances which cause inaccurate results. One of the well-known disadvantages of thermodilution is the overestimation of results at low values of cardiac output. This effect may concern the limited slew rate of the thermoelement mounted at the tip of the catheter. In this paper the relationship between the dynamic response of the thermoelement and the uncertainty of cardiac output measurements by means of thermodilution has been investigated theoretically and experimentally.
Źródło:
Metrology and Measurement Systems; 2012, 19, 4; 751-758
0860-8229
Pojawia się w:
Metrology and Measurement Systems
Dostawca treści:
Biblioteka Nauki
Artykuł
    Wyświetlanie 1-2 z 2

    Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies