- Tytuł:
- Calibration consistency of AC Voltage Standards developed at the Silesian University of Technology
- Autorzy:
-
Grzenik, M.
Kampik, M. - Powiązania:
- https://bibliotekanauki.pl/articles/114452.pdf
- Data publikacji:
- 2015
- Wydawca:
- Stowarzyszenie Inżynierów i Techników Mechaników Polskich
- Tematy:
-
AC-DC transfer
AC voltage standards
consistency of standards - Opis:
- In 2014 at the Silesian University of Technology (SUT) two calculable thermal AC voltage standards were developed. AC-DC differences of those standards were calculated from an appropriate mathematical model in 1 kHz – 1 MHz frequency range and determined experimentally in 10 Hz – 1 kHz frequency range. The correctness of determination of the AC-DC differences of the standards were verified by comparing them with a thermal converter calibrated in an external laboratory. That calibration confirmed the correctness of the determined AC-DC transfer differences. Lately, a comparison with a third laboratory has been performed. The paper presents results of these comparisons and a discovered inconsistency between the calculated AC-DC transfer differences of the SUT calculable thermal AC voltage standards and the AC-DC transfer difference of the AC voltage standard from the third laboratory.
- Źródło:
-
Measurement Automation Monitoring; 2015, 61, 5; 124-126
2450-2855 - Pojawia się w:
- Measurement Automation Monitoring
- Dostawca treści:
- Biblioteka Nauki