Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Wyszukujesz frazę "Gawlikowski, M." wg kryterium: Autor


Wyświetlanie 1-3 z 3
Tytuł:
Analysis of actual accuracy in cardiac output measurements by means of thermodilution
Autorzy:
Gawlikowski, M.
Pustelny, T.
Powiązania:
https://bibliotekanauki.pl/articles/201601.pdf
Data publikacji:
2012
Wydawca:
Polska Akademia Nauk. Czytelnia Czasopism PAN
Tematy:
cardiac output
hemodynamics monitoring
thermodilution
Opis:
The essential examination in hemodynamic monitoring of the patient is the cardiac output (CO) measurement. Nowadays, in clinical practice the most popular method is indicator dilution, particularly thermodilution. It is realized by applying the Swan-Ganz catheter and observing changes of the indicator concentration. This method is sensitive to many factors and the obtained results should be treated cautiously. The paper presents theoretical and experimental studies of metrological phenomena in CO measurements by means of thermodilution, paying special attention to medical aspects of the measurements. It has been pointed out that the actual unreliability of the mentioned method reaches values from 20% to 45%, which is in opposition to the technical data of patient's monitors (typical accuracy about 5%).
Źródło:
Bulletin of the Polish Academy of Sciences. Technical Sciences; 2012, 60, 3; 581-587
0239-7528
Pojawia się w:
Bulletin of the Polish Academy of Sciences. Technical Sciences
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
Theoretical and model analysis of the unreliability of cardiac output measurement by means of the thermodilution method
Autorzy:
Gawlikowski, M.
Pustelny, T.
Przywara-Chowaniec, B.
Nowak-Gawlikowska, J.
Powiązania:
https://bibliotekanauki.pl/articles/200095.pdf
Data publikacji:
2011
Wydawca:
Polska Akademia Nauk. Czytelnia Czasopism PAN
Tematy:
cardiac output measurement
thermodilution method
Opis:
Thermodilution is the clinically most often applied method of cardiac output measurements. This method is based on thermal indicator (iced isotonic salt solution) variation measurements by a Swan-Ganz catheter located inside the pulmonary artery. The unreliability of thermodilution should be estimated theoretically because of the lack of references. In this paper an attempt has been made to estimate theoretically the unreliability of thermodilution cardiac output measurements.
Źródło:
Bulletin of the Polish Academy of Sciences. Technical Sciences; 2011, 59, 4; 435-439
0239-7528
Pojawia się w:
Bulletin of the Polish Academy of Sciences. Technical Sciences
Dostawca treści:
Biblioteka Nauki
Artykuł
Tytuł:
Noninvasive acoustic blood volume measurement system for the POLVAD prosthesis
Autorzy:
Opilski, Z.
Konieczny, G.
Pustelny, T.
Gacek, A.
Kustosz, R.
Gawlikowski, M.
Powiązania:
https://bibliotekanauki.pl/articles/200093.pdf
Data publikacji:
2011
Wydawca:
Polska Akademia Nauk. Czytelnia Czasopism PAN
Tematy:
heart prosthesis
medical acoustics
blood volume measurement system
Opis:
The following paper presents researches concerning a noninvasive real-time blood volume measurement system applied in POLVAD prosthesis. The system is based on the acoustic Helmholtz resonator principle. The basis of the measurement method, followed by the preliminary tests of the possibility of incorporating the Helmholtz resonance idea into the POLVAD prosthesis is shown. The paper includes the actual measurement system construction and test results, both static and dynamic obtained at the Foundation for Cardiac Surgery Development in Zabrze, Poland. Conclusions and future plans are presented too.
Źródło:
Bulletin of the Polish Academy of Sciences. Technical Sciences; 2011, 59, 4; 429-433
0239-7528
Pojawia się w:
Bulletin of the Polish Academy of Sciences. Technical Sciences
Dostawca treści:
Biblioteka Nauki
Artykuł
    Wyświetlanie 1-3 z 3

    Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies