- Tytuł:
- Time-of-Flight Positron Emission Tomography with Resistive Plate Chamber Detectors: An Unlikely but Promising Approach
- Autorzy:
-
Couceiro, M.
Crespo, P.
Blanco, A.
Ferreira, N.
Mendes, L.
Ferreira Marques, R.
Fonte, P. - Powiązania:
- https://bibliotekanauki.pl/articles/1402607.pdf
- Data publikacji:
- 2015-05
- Wydawca:
- Polska Akademia Nauk. Instytut Fizyki PAN
- Tematy:
-
87.57.uk
29.40.Cs
87.57.cf
87.57.cm - Opis:
- The cost-effectiveness of resistive plate chamber detectors and their very good timing characteristics, open the possibility to build affordable time-of-flight positron emission tomography systems with a large axial field-of-view. Simulations suggest that, under reasonable assumptions, the absolute 3D true sensitivity, spatial resolution, and noise equivalent count rate of such systems for human whole-body screening, may exceed that of present crystal-based PET technology. However, due to the lack of energy resolution, although having energy sensitivity, the scatter fraction is expected to be considerably higher than that presented by crystal-based PET scanners. In the present paper, the simulation work done so far to access the expected performance of a resistive plate chamber time-of-flight-PET system with 2400 mm length axial field-of-view, a time resolution of 300 ps full width at half maximum for photons pairs, and depth-of-interaction information, will be revised.
- Źródło:
-
Acta Physica Polonica A; 2015, 127, 5; 1453-1461
0587-4246
1898-794X - Pojawia się w:
- Acta Physica Polonica A
- Dostawca treści:
- Biblioteka Nauki