This paper presents the study of algorithms for derivation of respiration waveform from the electrocardiogram. The problem has considerable clinical impact, because the heart rate and respiration are both driven by the central nervous system, and commonly used low-cost Holter recording may be used for efficient detection of breath disturbances (e.g. apnea). Three methods based on: heart rate, heart position and lung resistance influencing the ECG amplitude were compared in our research. Among 18 volunteers breathing at a controlled frequency all implemented algorithms show acceptable sensitivity of order of 97% in slow breathing phases detection. In fast breathing the sensitivity is reduced to 90%, since the heart beats are too sparse with regard of respiration waveform.
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies