Optimization of ultra-performance liquid chromatography (UPLC) with fluorescence detector (FLD) method for the quantitative determination of selected neurotransmitters in rat brain
Optimization of ultra-performance liquid chromatography (UPLC) with fluorescence detector (FLD) method for the quantitative determination of selected neurotransmitters in rat brain
Background
Glutamate (Glu) and γ-aminobutyric acid (GABA) are the main neurotransmitters in the central nervous system for excitatory and inhibitory processes, respectively. Monitoring these neurotransmitters is an essential tool in establishing pathological functions, among others in terms of occupational exposure to toxic substances.
Material and Methods
We present modification of the HPLC (high-performance liquid chromatography) to the UPLC (ultra-performance liquid chromatography) method for the simultaneous determination of glutamate and γ-aminobutyric acid in a single injection. The isocratic separation of these neurotransmitter derivatives was performed on Waters Acquity BEH (ethylene bridged hybrid) C18 column with particle size of 1.7 μm at 35°C using a mobile phase consisting of 0.1 M acetate buffer (pH 6.0) and methanol (60:40, v/v) at a flow rate of 0.3 ml/min. The analytes were detected with the fluorescence detector (FLD) using derivatization with o-phthaldialdehyde (OPA), resulting in excitation at 340 nm and emission at 455 nm.
Results
Several validation parameters including linearity (0.999), accuracy (101.1%), intra-day precision (1.52–1.84%), inter-day precision (2.47–3.12%), limit of detection (5–30 ng/ml) and quantification (100 ng/ml) were examined. The developed method was also used for the determination of these neurotransmitters in homogenates of selected rat brain structures.
Conclusions
The presented UPLC-FLD is characterized by shorter separation time (3.5 min), which is an adaptation of the similar HPLC methods and is an alternative for more expensive references techniques such as liquid chromatography coupled with tandem mass-spectrometry (LC-MS/MS) methods. Med Pr 2017;68(5):583–591
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies