A BaF$\text{}_{2}$ positron-lifetime spectrometer equipped with a coincidence system that represents a compromise between the traditional fast-fast and fast-slow arrangements is described. The main difference of the present configuration from both the fast-slow and fast-fast ones consists in use of a sum of the energy signals from the start and stop detectors to select the coincidence events. Quality of the spectrometer response function (≈ 150 ps FWHM for $\text{}^{22}$Na) is very close to that observed for our fast-slow configuration with the equivalent detectors but throughput of the present apparatus is increased almost by a factor of two. Moreover, the electronic scheme becomes less complicated, which has also a positive impact on its cost.
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies