Measurements of α-particle sources require corrections to the counting rate due to scattering and self-
-absorption in the source and the backing material. In this study, we describe a simple procedure to estimate these
corrections using the new Monte Carlo code AlfaMC to consider the effects of scattering and self-absorption
conjointly, and so to determine the activity of α emitters. The procedure proposed was applied to 235UO2 sources
deposited on highly polished platinum backings. In general, the dependence of the efficiency with source thickness
was in good agreement with a simple model considering a linear and a hyperbolic behavior for thin and
thick sources, respectively, although significant deviations from this model were found for very thin sources.
For these very thin sources, the Monte Carlo simulation revealed to be as a required method in the primary
calibration of α-particle sources. The efficiency results obtained by simulation with AlfaMC were in agreement
with available efficiency data.
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies