This article is about realization of the laboratory stand used for measuring parameters of LCD displays. The stand together with additional accessories (spectrophotometer and measuring probes) as well as software allow to measure a wide range of parameters (colour gamut, response time, contrast coefficient and its irregularity, luminance of black colour and its irregularity, luminance of white colour and its irregularity, changes of luminance depending on the viewing angle). This paper also show method and results of measurements of differences in contrast ratio and luminance uniformity between particular pieces of the same model of a display screen (research has been carried out on 5 pieces of the same model of a display). The presented results show that the measured value of the viewing angle may significantly differ from the value provided by producers of displays.
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies