Measurement of pCO2 in highly dynamic coastal zones such as the southern Baltic
Sea presents many challenges. In this study, we designed a system to measure pCO2 and then
validated it in a series of laboratory and seagoing tests. The fast response time of the system
was shown to provide a better resolution of CO2 system gradients. In the open waters of the
Baltic Sea, the accuracy of the pCO2 measurements (±1.3 μatm) met the requirements of the
ICOS (±2.0 μatm). In the coastal zone, there was less consistency between pCO2, DIC and pH
measurements, suggesting the need to redefine the quality assurance and control requirements
for the measurement of pCO2 in dynamic regions.
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies