The paper presents the results of investigations concerning the noninvasive method of estimating
the actual volume of the blood chamber of the POLVAD-EXT type ventricular assist device (VAD)
during its operation. The proposed method is based on the principle of Helmholtz’s acoustic resonance.
Both the theory, main stages of the development of the measurement method as well as the practical
implementation of the proposed method in the physical model of the POLVAD-EXT device are dealt with.
The paper contains the results of static measurements by means of the proposed method (conducted at the
Department of Optoelectronics, Silesian University of Technology) as well as the dynamic measurements
taken at the Foundation of Cardiac Surgery Development (Zabrze, Poland) with the professional model
of the human cardiovascular system. The results of these measurements prove that the proposed method
allows to estimate the actual blood chamber volume with uncertainties below 10%.
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies