In this study after benchmarking of Monte Carlo (MC) simulation of a 6 MV linac, the simulation model was
used for estimation of tumor dose enhancement by gold nanoparticles (GNPs). The 6 MV photon mode of a Siemens
Primus linac was simulated and a percent depth dose and dose profiles values obtained from the simulations were compared
with the corresponding measured values. Dose enhancement for various sizes and concentrations of GNPs were
studied for two cases with and without the presence of a flattening filter in the beam’s path. Tumor dose enhancement
with and without the presence of the flattening filter was, respectively 1–5 and 3–10%. The maximum dose enhancement
was observed when 200 nm GNPs was used and the concentration was 36 mg/g tumor. Furthermore, larger GNPs
resulted in higher dose values in the tumor. After careful observation of the dose enhancement factor data, it was found
that there is a poor relation between the nanoparticle size and dose enhancement. It seems that for high energy photons,
the dose enhancement is more affected by the concentration of nanoparticles than their size.
Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies